pH sensorOPTISENS PH 9100Handbook© KROHNE 09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 enThe documentation is only complete when used in combination
2 DEVICE DESCRIPTION 10 OPTISENS PH 9100www.krohne.com 09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 enThe OPTISENS PH 9100 is manufactured using a p
INSTALLATION 311OPTISENS PH 9100www.krohne.com09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en3.1 Notes on installation3.2 Storage and transport•
3 INSTALLATION 12 OPTISENS PH 9100www.krohne.com 09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 enStoring the sensor in the provided plastic tube• Sc
INSTALLATION 313OPTISENS PH 9100www.krohne.com09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en3.4 Pre-installation requirementsUnpacking the sensor
3 INSTALLATION 14 OPTISENS PH 9100www.krohne.com 09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en3.5.1 Connecting the sensor cable to the signal con
INSTALLATION 315OPTISENS PH 9100www.krohne.com09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 enThe following instructions describe the connection of
3 INSTALLATION 16 OPTISENS PH 9100www.krohne.com 09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en3.5.3 Connecting the cable to the sensorConnecting
INSTALLATION 317OPTISENS PH 9100www.krohne.com09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en3.7 Installing the sensor 3.7.1 General installation
3 INSTALLATION 18 OPTISENS PH 9100www.krohne.com 09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en3.7.2 Mounting to a flow-through holderInstalling a
INSTALLATION 319OPTISENS PH 9100www.krohne.com09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en3.8 Installing an external temperature sensorSince th
All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without the prior written authorisation of KROHNE Messtech
3 INSTALLATION 20 OPTISENS PH 9100www.krohne.com 09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en3.9 Examples of a typical measuring pointThe follow
OPERATION 421OPTISENS PH 9100www.krohne.com09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en4.1 Menu mode structureINFORMATION!The following table j
4 OPERATION 22 OPTISENS PH 9100www.krohne.com 09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en 3 or 4 pages, scrolling with ↓ or ↑>2.5s^B test >
OPERATION 423OPTISENS PH 9100www.krohne.com09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en4.2 Function tables4.2.1 Menu A, quick setupA9, ph cal.
4 OPERATION 24 OPTISENS PH 9100www.krohne.com 09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en4.2.2 Menu B, testThe procedure to start the simulatio
OPERATION 425OPTISENS PH 9100www.krohne.com09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en4.2.3 Menu C, setupC1, process input AC2, process input
4 OPERATION 26 OPTISENS PH 9100www.krohne.com 09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 enC1.15.5C2.15.5temp. comp. Menu for activating the tempe
OPERATION 427OPTISENS PH 9100www.krohne.com09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en4.3 Calibration4.3.1 Temperature compensationThere are
4 OPERATION 28 OPTISENS PH 9100www.krohne.com 09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 enAfter starting-up the signal converter, the measuring s
OPERATION 429OPTISENS PH 9100www.krohne.com09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 enStep 2: Configure /adjust the temperature sensor• If nec
CONTENTS 3www.krohne.com09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 enOPTISENS PH 91001 Safety instructions 51.1 Intended use ...
4 OPERATION 30 OPTISENS PH 9100www.krohne.com 09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en• Enter the measured temperature. Press ^ to confirm t
OPERATION 431OPTISENS PH 9100www.krohne.com09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en4.3.2 Calibrating pH measurementA pH calibration is nece
4 OPERATION 32 OPTISENS PH 9100www.krohne.com 09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 enStep 1: activating the manual hold function• You have
OPERATION 433OPTISENS PH 9100www.krohne.com09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 enStep 3a: accessing the calibration menu via the main menu
4 OPERATION 34 OPTISENS PH 9100www.krohne.com 09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 enStep 4: calibration procedure• After choosing the subm
OPERATION 435OPTISENS PH 9100www.krohne.com09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en• Press ^ to confirm.i The slope of the sensor is displa
4 OPERATION 36 OPTISENS PH 9100www.krohne.com 09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en4.3.3 Calibration logAccessing the calibtration log•
SERVICE 537OPTISENS PH 9100www.krohne.com09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en5.1 Maintenance5.1.1 Cleaning• Slight dirt residues or d
5 SERVICE 38 OPTISENS PH 9100www.krohne.com 09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en5.2 Spare parts availabilityThe manufacturer adheres to
SERVICE 539OPTISENS PH 9100www.krohne.com09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en5.4.2 Form (for copying) to accompany a returned device5.5
CONTENTS 4 www.krohne.com 09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 enOPTISENS PH 91005.1 Maintenance ...
6 TECHNICAL DATA 40 OPTISENS PH 9100www.krohne.com 09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en6.1 Measuring principle6.1.1 pH measurementThe m
TECHNICAL DATA 641OPTISENS PH 9100www.krohne.com09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 enThe voltage change of a pH sensor at 25°C/ 77°F is a
6 TECHNICAL DATA 42 OPTISENS PH 9100www.krohne.com 09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en6.2 Technical datapH sensorINFORMATION!•The follo
TECHNICAL DATA 643OPTISENS PH 9100www.krohne.com09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 enOperating conditionsTemperature range -5…+100°C / +2
6 TECHNICAL DATA 44 OPTISENS PH 9100www.krohne.com 09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en6.3 DimensionsFigure 6-4: Dimension of the pH sen
APPENDIX 745OPTISENS PH 9100www.krohne.com09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en7.1 pH as a function of mVThe pH-value is the negative de
7 APPENDIX 46 OPTISENS PH 9100www.krohne.com 09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en7.2 pH temperature dependencyThe output from a pH-senso
847OPTISENS PH 9100www.krohne.com09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en
KROHNE product overview• Electromagnetic flowmeters• Variable area flowmeters• Ultrasonic flowmeters• Mass flowmeters• Vortex flowmeters• Flow control
SAFETY INSTRUCTIONS 15OPTISENS PH 9100www.krohne.com09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en1.1 Intended useThe intended use of OPTISENS PH
1 SAFETY INSTRUCTIONS 6 OPTISENS PH 9100www.krohne.com 09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en1.2.2 DisclaimerThe manufacturer will not be
SAFETY INSTRUCTIONS 17OPTISENS PH 9100www.krohne.com09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en1.2.5 Warnings and symbols usedSafety warnings
2 DEVICE DESCRIPTION 8 OPTISENS PH 9100www.krohne.com 09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en2.1 Scope of deliveryOptional accessories• SEN
DEVICE DESCRIPTION 29OPTISENS PH 9100www.krohne.com09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en2.2 Device description2.2.1 pH sensorFigure 2-2
Comments to this Manuals