ALTOSONIC V 5-beam ultrasonic flowmetering system operating and installation instructions UFC-V/…-EEx Flow Converter UFS 500 F/5STR-EEx
10 www.krohne.com 04/2012 ALTOSONIC V MECHANICAL INSTALLATION2 2. Mechanical installation 2.1 Ultrasonic Flow Sensor (UFS-V) Handling Check wei
04/2012 www.krohne.com 11 ALTOSONIC V MECHANICAL INSTALLATION2Do not expose the sensor unit to intensive vibrations. Support of the flow sensor i
12 www.krohne.com 04/2012 ALTOSONIC V MECHANICAL INSTALLATION2 are pre-wired. At one side cable glands are mounted. At each corner outside the hou
04/2012 www.krohne.com 13 ALTOSONIC V ELECTRICAL INSTALLATION33. Electrical installation 3.1 Overview of connections The figure below shows an o
14 www.krohne.com 04/2012 ALTOSONIC V ELECTRICAL INSTALLATION3 It is good practice to use shielded twisted pair solutions for I/O wiring. R
04/2012 www.krohne.com 15 ALTOSONIC V ELECTRICAL INSTALLATION3Mains connection of the UFC-V (hazardous environment) Use a 3 wire power cable w
16 www.krohne.com 04/2012 ALTOSONIC V ELECTRICAL INSTALLATION3 Connection box on Flow Sensor UFS-V and shield grounding clamp On the side of
04/2012 www.krohne.com 17 ALTOSONIC V ELECTRICAL INSTALLATION3 Connector X1 + X2 of Flow Converter UFC-V 24 VDC Connector X1 + X2 of Flow Conve
18 www.krohne.com 04/2012 ALTOSONIC V ELECTRICAL INSTALLATION3 3.5 Connecting Flow Sensor to the Flow Computer The Flow Sensor UFS-V also has to b
04/2012 www.krohne.com 19 ALTOSONIC V ELECTRICAL INSTALLATION33.6 Connecting the Flow Converter to the Flow Computer Between the Flow Converter U
2 www.krohne.com 04/2012 ALTOSONIC V General advice on safety Do not install, operate or maintain this flowmeter without reading, understandi
20 www.krohne.com 04/2012 ALTOSONIC V ELECTRICAL INSTALLATION3 Connector X1 I/O-rack Flow Computer UFP-V Attention this is an example, posi
04/2012 www.krohne.com 21 ALTOSONIC V ELECTRICAL INSTALLATION33.8 Connecting the outputs of the UFP-V I/O-rack The I/O signals are wired to the c
22 www.krohne.com 04/2012 ALTOSONIC V ELECTRICAL INSTALLATION3 Connection parameters per type for UFP card MP103 Analog inputs AD812 card PCLD
04/2012 www.krohne.com 23 ALTOSONIC V ELECTRICAL INSTALLATION33.9 Computer and I/O rack The custody certified Flow Processor UFP-V and I/O rack a
24 www.krohne.com 04/2012 ALTOSONIC V ELECTRICAL INSTALLATION3 Construction The Flow Processor UFP-V, is an industrial type computer, e.g. Advante
04/2012 www.krohne.com 25 ALTOSONIC V ELECTRICAL INSTALLATION3 Top view I/O module MTL 5073 UFS-V junction box detail Field cabling For the
26 www.krohne.com 04/2012 ALTOSONIC V ELECTRICAL INSTALLATION3 Example of a configuration with interfacing modules (front side) Connections ter
04/2012 www.krohne.com 27 ALTOSONIC V ELECTRICAL INSTALLATION3Rear of the I/O rack Items marked with connections to the UFP-V Flow Processor are
28 www.krohne.com 04/2012 ALTOSONIC V ELECTRICAL INSTALLATION3 Side view Advantec Workstation AWS-842T 3.10 Modbus connection/communication to e
04/2012 www.krohne.com 29 ALTOSONIC V SERVICING44. Returning the flowmeter for service or repair Your ultrasonic flowmeter has been carefully man
04/2012 www.krohne.com 3 ALTOSONIC V 1Disclaimer This document contains important information on the instrument. KROHNE attempts to be as accurat
30 www.krohne.com 04/2012 ALTOSONIC V SERVICING 4 4.2 Spare parts For more information please contact KROHNE!
KROHNE product overview• Electromagnetic flowmeters• Variable area flowmeters• Ultrasonic flowmeters• Mass flowmeters• Vortex flowmeters• Flow control
4 www.krohne.com 04/2012 ALTOSONIC V Documentation supplied Operating and installation instructions Approval documents, unless reproduced
04/2012 www.krohne.com 5 ALTOSONIC V 1Table of Contents 1. Introduction ...
6 www.krohne.com 04/2012 ALTOSONIC V INTRODUCTION 1 1. Introduction 1.1 Cautions Refer all maintenance or service to trained KROHNE service engine
04/2012 www.krohne.com 7 ALTOSONIC V INTRODUCTION11.3 System description The ALTOSONIC V is a high precision instrument designed for linear, bi
8 www.krohne.com 04/2012 ALTOSONIC V INTRODUCTION 1 3 wire shielded wire twisted (or two pairs) Power supply to the I/O-rack: 3 wire includi
04/2012 www.krohne.com 9 ALTOSONIC V INTRODUCTION1 The UFP-V Flow Processor and I/O rack are designed to operate safely under the following condi