Installation and
operating instructions
IFC 110 F V2.0
IFC 110 F-EEx V2.0
Signal converters for
electromagnetic flowmeters
Applicable to
Software Versions
Display & control unit
No. 3.19937.02.00
A/D converter
No. 8.13393.02.00
Outputs/inputs (I/O)
No. 3.16230.01.00
How to use these Instructions
Flowmeters are delivered ready for operation.
The flow sensor must be installed in the pipeline as described in the instructions for
installation inside the packing of the flow sensor.
Connection of power supply (Sect. 1.1-1.2) Pages 4- 5
Electrical connection between IFC 110 F and primary head (Sect. 1.3) Pages 6-13
Electrical connection of outputs and inputs (Sect. 2) Pages 14-22
Start-up (Sect. 3) Pages 23
Power the flowmeter. THAT’S ALL. The system is operative.
Operator control of the IFC 110 F signal converter is described in Sect. 4.
Variable area flowmeters
Vortex flowmeters
Flow controllers
Electromagnetic flowmeters
Ultrasonic flowmeters
Mass flowmeters
Level measuring instruments
Communications technology
Engineering systems & solutions
Switches, counters, displays and recorders
Heat metering
Pressure and temperature
DIN A4: 7.02264.31.00
©
KROHNE 08/2003 US size: 7.02264.71.00
Page view 0
1 2 3 4 5 6 ... 42 43

Summary of Contents

Page 1 - IFC 110 F-EEx V2.0

Installation and operating instructionsIFC 110 F V2.0IFC 110 F-EEx V2.0Signal converters for electromagnetic flowmetersApplicable toSoftware Versions●

Page 2 - Instrument nameplates

10 IFC 110 F 05/2003 1.3.5 Connection diagrams for power supply and flow sensors Important remarks for circuit diagrams PLEASE NOTE ! •

Page 3 - Software history

05/2003 IFC 110 F 11 Process temperature below 150°C (302°F) I Signal cable A (type DS) II Signal cable B (type BTS) IFC 110 F

Page 4 - 4 IFC 110 F

12 IFC 110 F 05/2003 1.3.6 EEx-Connection diagrams for power supply and flow sensors PLEASE NOTE! • The figures in brackets indicate the stra

Page 5 - IFC 110 F 5

05/2003 IFC 110 F 13 Process temperature below 150°C (302°F) I Signal cable A (type DS) II Signal cable B (type BTS) IFC 110 F

Page 6 - 6 IFC 110 F

14 IFC 110 F 05/2003 2 Electrical connection: outputs and inputs IMPORTANT! For EEx versions, also pay regard to all directions included in Se

Page 7 - IFC 110 F 7

05/2003 IFC 110 F 15 2.2 Current output I • The current output is electrically isolated from all other circuits. • Setting data and functio

Page 8 - 8 IFC 110 F

16 IFC 110 F 05/2003 2.3 Pulse outputs P and A1 2.3.1 Pulse output P for electronic totalizers (EC) • Pulse output P is electrically isolate

Page 9 - IFC 110 F 9

05/2003 IFC 110 F 17 2.3.2 Pulse output A1 for electromechanical totalizers (EMC) PLEASE NOTE: The output terminal A1 can be used as status o

Page 10 - 10 IFC 110 F

18 IFC 110 F 05/2003 2.4 Status outputs A1 / A2 / D1 / D2 PLEASE NOTE: The output terminal A1 can be used as status output A1 or as a 2nd pulse

Page 11 - IFC 110 F 11

05/2003 IFC 110 F 19 2.6 Connection diagrams of outputs and inputs IMPORTANT! For EEx versions, also pay regard to all directions included in

Page 12 - 12 IFC 110 F

2 IFC 110 F 05/2003 Signal converter versions The operating data are factory-set to your ordered specifications. IFC 110 F / D Standard versio

Page 13 - IFC 110 F 13

20 IFC 110 F 05/2003 { Current output Iactive | Current output Iactive with automatic range change BA without external change-over relay }

Page 14 - 14 IFC 110 F

05/2003 IFC 110 F 21  Pulse output Pactive  Pulse output Pactive for electronic totalizers (EC) for frequencies ≤ 1 kHz for electronic tot

Page 15 - 2.2 Current output I

22 IFC 110 F 05/2003  Status outputs D1 / D2 / A1 / A2 active Status outputs D1 / D2 / A1 / A2 passive I ≤ 100 mA Uext. ≤ 32 V DC / ≤ 24 V AC

Page 16 - 2.3 Pulse outputs P and A1

05/2003 IFC 110 F 23 3 Start-up • Before connecting to power, check that the instrument is correctly installed as described in Sections 1 and

Page 17 - IFC 110 F 17

24 IFC 110 F 05/2003 4 Operating of the signal converter 4.1 KROHNE operating concept

Page 18 - 18 IFC 110 F

05/2003 IFC 110 F 25 4.2 Operating and control elements The instrument can be operated by means of ... … the 1

Page 19 - IFC 110 F 19

26 IFC 110 F 05/2003 4.3 Key functions In the following, the cursor or flashing part of the display is shown against a grey background. To star

Page 20 - 20 IFC 110 F

05/2003 IFC 110 F 27 To change numbers increase number 3 9 7. 3 5 ↑

Page 21 - IFC 110 F 21

28 IFC 110 F 05/2003 4.4 Table of settable functions Abbreviations used: A1, A2 Status outputs P (P2) Pulse output (2nd pulse output A1) (A1

Page 22 - 22 IFC 110 F

05/2003 IFC 110 F 29 Fct. Text Description and setting 1.04 DISPLAY Display functions → DISP.FLOW Selection of flow display ● NO DISP. ●

Page 23 - 3 Start-up

05/2003 IFC 110 F 3 System description Electromagnetic flowmeters are precision instruments designed for linear flow measurement of liquid prod

Page 24 - 24 IFC 110 F

30 IFC 110 F 05/2003 Fct. Text Description and setting 1.08 STATUS A2 Status outputs A2, D1 and D2 1.09 STATUS D1 Description of function of

Page 25 - IFC 110 F 25

05/2003 IFC 110 F 31 Fct. Text Description and setting 1.07 STATUS A1 Status output A1 (terminal A1 connected as status output A1 or as a 2nd

Page 26 - 4.3 Key functions

32 IFC 110 F 05/2003 Fct. Text Description and setting 3.0 INSTALL. Installation menu 3.01 LANGUAGE Select language for display texts ●

Page 27 - IFC 110 F 27

05/2003 IFC 110 F 33 Fct. Text Description and setting 3.05 USER UNIT Set any required unit for flowrate and counting → TEXT VOL. Set text f

Page 28 - Abbreviations used:

34 IFC 110 F 05/2003 Fct. Text Description and setting → LIMIT VAL. Set limit value for noise/interference suppression (appears only when

Page 29

05/2003 IFC 110 F 35 4.6 Resetting the totalizer and deleting error messages, RESET/QUIT menu Delete error messages in RESET/QUIT menu Key

Page 30

36 IFC 110 F 05/2003 6 Special Applications, Functional Checks, Service and Order Numbers 6.1 Use in hazardous areas 6.1.1 General Signal con

Page 31 - X . X X X X X . X X

05/2003 IFC 110 F 37 6.1.3 Installation and electrical connection Type IFC 110 F - EEx signal converters are type tested as associated electrica

Page 32

38 IFC 110 F 05/2003 10 Technical data 10.1 Signal converter Mode of operation and system structure Measurement principle Faraday’s law of

Page 33

05/2003 IFC 110 F 39 Status outputs (passive) D1 / D2 / A2 A1 (can also be operated as pulse output) Function, set for trip point trip point

Page 34

4 IFC 110 F 05/2003 1 Electrical connection: power supply 1.1 Location and important installation notes ……………………. PLEASE NOTE ! • Electrical

Page 35

40 IFC 110 F 05/2003 10.2 Error limits Display, digital values, pulse output F maximum error in % of measured value (MV), not typical values v

Page 36 - 6.1 Use in hazardous areas

05/2003 IFC 110 F 41 10.3 Dimensions and weights IFC 110 F / IFC 110 F-EEx and ZD / ZD-EEx Dimensions in mm and inch IFC 110 F Sig

Page 37 - IFC 110 F 37

42 IFC 110 F 05/2003 13 Approvals 13.1 EC-type examination certificate English translation 13.2 EC-type examination certificate German origin

Page 38 - 10.1 Signal converter

05/2003 IFC 110 F 43 If you need to return flowmeters for testing or repair to KROHNE Your electromagnetic flowmeter • has been carefully manu

Page 39 - IFC 110 F 39

05/2003 IFC 110 F 5 1.2 Power supply - connection PLEASE NOTE ! • Type of enclosure IP 65 to IEC 529 / EN 60529 equivalent to NEMA 4/4X. •

Page 40 - 10.2 Error limits

6 IFC 110 F 05/2003 1.3 Electrical connection of flow sensors 1.3.1 General remarks on signal lines A and B and field current line C Proper o

Page 41 - 10.4 Flow table

05/2003 IFC 110 F 7 1.3.2 Stripping (preparation) of signal cables Please note: The numbers in the drawings designate the stranded drain wires

Page 42 - 13 Approvals

8 IFC 110 F 05/2003 1.3.3 Grounding of flow sensor • The flow sensor must be correctly connected to ground. • The grounding cable may not tra

Page 43 - IFC 110 F 43

05/2003 IFC 110 F 9 1.3.4 Cable lengths (max. distance between signal converter and flow sensor) Abbreviations and explanations The abbreviatio