pH sensorOPTISENS PH 8100OPTISENS PH 8100OPTISENS PH 8100OPTISENS PH 8100HandbookHandbookHandbookHandbook© KROHNE 03/2012 - 4001925901 MA OPTISENS PH
2 DEVICE DESCRIPTION 10 OPTISENS PH 8100www.krohne.com 03/2012 - 4001925901 MA OPTISENS PH 8100 R01 enThe OPTISENS PH 8100 is manufactured using a pH
INSTALLATION 311OPTISENS PH 8100www.krohne.com03/2012 - 4001925901 MA OPTISENS PH 8100 R01 enInstallation3.1 Notes on installation3.2 Storage and t
3 INSTALLATION 12 OPTISENS PH 8100www.krohne.com 03/2012 - 4001925901 MA OPTISENS PH 8100 R01 enStoring the electrode in the provided plastic tube• S
INSTALLATION 313OPTISENS PH 8100www.krohne.com03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en3.4 Pre-installation requirementsUnpacking the electrod
3 INSTALLATION 14 OPTISENS PH 8100www.krohne.com 03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en3.5.1 Connecting the sensor cable to the signal conve
INSTALLATION 315OPTISENS PH 8100www.krohne.com03/2012 - 4001925901 MA OPTISENS PH 8100 R01 enThe following instructions describe the connection of th
3 INSTALLATION 16 OPTISENS PH 8100www.krohne.com 03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en3.5.3 Connecting the cable to the sensorConnecting th
INSTALLATION 317OPTISENS PH 8100www.krohne.com03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en3.7 Installing the sensor 3.7.1 General installation i
3 INSTALLATION 18 OPTISENS PH 8100www.krohne.com 03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en3.7.2 Mounting to a flow-through holderInstalling a n
INSTALLATION 319OPTISENS PH 8100www.krohne.com03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en3.7.3 Mounting to SENSOFIT IMM 1000 immersion holderINF
All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without the prior written authorisation of KROHNE Messtech
3 INSTALLATION 20 OPTISENS PH 8100www.krohne.com 03/2012 - 4001925901 MA OPTISENS PH 8100 R01 enInstalling a new electrode (step 1)• Pull the cap off
INSTALLATION 321OPTISENS PH 8100www.krohne.com03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en3.8 Installing an external temperature sensorSince the
3 INSTALLATION 22 OPTISENS PH 8100www.krohne.com 03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en3.9 Examples of a typical measuring pointThe followin
OPERATION 423OPTISENS PH 8100www.krohne.com03/2012 - 4001925901 MA OPTISENS PH 8100 R01 enOperation4.1 Menu mode structureINFORMATION!The following
4 OPERATION 24 OPTISENS PH 8100www.krohne.com 03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en 3 or 4 pages, scrolling with ↓ or ↑>2.5s^B testB test
OPERATION 425OPTISENS PH 8100www.krohne.com03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en4.2 Function tables4.2.1 Menu A, quick setupA9, ph cal. B
4 OPERATION 26 OPTISENS PH 8100www.krohne.com 03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en4.2.2 Menu B, testThe procedure to start the simulation
OPERATION 427OPTISENS PH 8100www.krohne.com03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en4.2.3 Menu C, setupC1, process input AC1, process input AC
4 OPERATION 28 OPTISENS PH 8100www.krohne.com 03/2012 - 4001925901 MA OPTISENS PH 8100 R01 enC1.15.5C2.15.5temp. comp. Menu for activating the tempera
OPERATION 429OPTISENS PH 8100www.krohne.com03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en4.3 Calibration4.3.1 Temperature compensationThere are th
CONTENTS 3www.krohne.com03/2012 - 4001925901 MA OPTISENS PH 8100 R01 enOPTISENS PH 81001 Safety instructions 51.1 Intended use ...
4 OPERATION 30 OPTISENS PH 8100www.krohne.com 03/2012 - 4001925901 MA OPTISENS PH 8100 R01 enAfter starting-up the signal converter, the measuring scr
OPERATION 431OPTISENS PH 8100www.krohne.com03/2012 - 4001925901 MA OPTISENS PH 8100 R01 enStep 2: Configure /adjust the temperature sensor• If neces
4 OPERATION 32 OPTISENS PH 8100www.krohne.com 03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en• Enter the measured temperature. Press ^ to confirm the
OPERATION 433OPTISENS PH 8100www.krohne.com03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en4.3.2 Calibrating pH measurementA pH calibration is necess
4 OPERATION 34 OPTISENS PH 8100www.krohne.com 03/2012 - 4001925901 MA OPTISENS PH 8100 R01 enStep 1: activating the manual hold function• You have ac
OPERATION 435OPTISENS PH 8100www.krohne.com03/2012 - 4001925901 MA OPTISENS PH 8100 R01 enStep 3a: accessing the calibration menu via the main menu s
4 OPERATION 36 OPTISENS PH 8100www.krohne.com 03/2012 - 4001925901 MA OPTISENS PH 8100 R01 enStep 4: calibration procedure• After choosing the submen
OPERATION 437OPTISENS PH 8100www.krohne.com03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en• Press ^^^^ to confirm.i The slope of the electrode is di
4 OPERATION 38 OPTISENS PH 8100www.krohne.com 03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en4.3.3 Calibration logAccessing the calibtration log• Yo
SERVICE 539OPTISENS PH 8100www.krohne.com03/2012 - 4001925901 MA OPTISENS PH 8100 R01 enService5.1 Maintenance5.1.1 Cleaning• Slight dirt residues
CONTENTS 4 www.krohne.com 03/2012 - 4001925901 MA OPTISENS PH 8100 R01 enOPTISENS PH 81005 Service 395.1 Maintenance ...
5 SERVICE 40 OPTISENS PH 8100www.krohne.com 03/2012 - 4001925901 MA OPTISENS PH 8100 R01 enThe following figure shows the aging effects:5.2 Spare par
SERVICE 541OPTISENS PH 8100www.krohne.com03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en5.4 Returning the device to the manufacturer5.4.1 General i
5 SERVICE 42 OPTISENS PH 8100www.krohne.com 03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en5.4.2 Form (for copying) to accompany a returned device5.5
TECHNICAL DATA 643OPTISENS PH 8100www.krohne.com03/2012 - 4001925901 MA OPTISENS PH 8100 R01 enTechnical data6.1 Measuring principle6.1.1 pH measur
6 TECHNICAL DATA 44 OPTISENS PH 8100www.krohne.com 03/2012 - 4001925901 MA OPTISENS PH 8100 R01 enThe voltage change of a pH electrode at 25°C/ 77°F i
TECHNICAL DATA 645OPTISENS PH 8100www.krohne.com03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en6.2 Technical dataINFORMATION!•The following data is
6 TECHNICAL DATA 46 OPTISENS PH 8100www.krohne.com 03/2012 - 4001925901 MA OPTISENS PH 8100 R01 enOperating conditionsTemperature range 0…+130°C / +32
TECHNICAL DATA 647OPTISENS PH 8100www.krohne.com03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en6.3 DimensionsFigure 6-4: Dimension of OPTISENS PH 81
6 TECHNICAL DATA 48 OPTISENS PH 8100www.krohne.com 03/2012 - 4001925901 MA OPTISENS PH 8100 R01 enFigure 6-6: Dimensions SENSOFIT IMM 1000Dimensions [
APPENDIX 749OPTISENS PH 8100www.krohne.com03/2012 - 4001925901 MA OPTISENS PH 8100 R01 enAppendix7.1 pH as a function of mVThe pH-value is the negat
SAFETY INSTRUCTIONS 15OPTISENS PH 8100www.krohne.com03/2012 - 4001925901 MA OPTISENS PH 8100 R01 enSafety instructions1.1 Intended useThe intended u
7 APPENDIX 50 OPTISENS PH 8100www.krohne.com 03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en7.2 pH temperature dependencyThe output from a pH-electro
NOTES 851OPTISENS PH 8100www.krohne.com03/2012 - 4001925901 MA OPTISENS PH 8100 R01 enNotes.book Page 51 Wednesday, July 4, 2012 11:07 AM
KROHNE product overview• Electromagnetic flowmeters• Variable area flowmeters• Ultrasonic flowmeters• Mass flowmeters• Vortex flowmeters• Flow control
1 SAFETY INSTRUCTIONS 6 OPTISENS PH 8100www.krohne.com 03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en1.2.2 DisclaimerThe manufacturer will not be li
SAFETY INSTRUCTIONS 17OPTISENS PH 8100www.krohne.com03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en1.2.5 Warnings and symbols usedSafety warnings ar
2 DEVICE DESCRIPTION 8 OPTISENS PH 8100www.krohne.com 03/2012 - 4001925901 MA OPTISENS PH 8100 R01 enDevice description2.1 Scope of deliveryOptional
DEVICE DESCRIPTION 29OPTISENS PH 8100www.krohne.com03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en2.2 Device description2.2.1 pH electrodesFigure 2