05/983.1M54EA1 059821 Order No. DIN size 7.02143.31.00US size 7.02143.71.000.73.00CONTENTSInstallation in the pipeline Pages 4-5 and 7-8Grounding Page
10Switch off power source before commencing work!1) Note down terminal assignment before dismantling the “old“ primary head.2) Install the new primary
1110 Technical dataSize/type DN 2.5 - 80 and 1/10” - 3”Available connections see page 3Electrical conductivity ≥ 5 µS/cm (≥ 20 µS/cm for deminerali
1211 Dimensions and weightsPLEASE NOTEThe total dimension for the height is obtained from dimension b (see table) plus the heightof the terminal bo
13160 (6.30,,)136 (5.35,,)098 (3.86,,)208 (8.19,,) Ex: 238 (9.37,,)140 (5.51,,)078 (3.07,,)Lm071 (2.80,,)105 (4.13,,)165 (6.50,,) ebdDimensions with
14Dairy screw connection to DIN 11 851Connection meter size dia. D LAaDN 10 1)038 (1.50)47.0 (1.85)214 ( 8.43)DN 15DN 25 063 (2.48) 230 (9.06)D
15If you need to return flowmeters for testingor repair to KrohneSPECIMEN certificateYour electromagnetic flowmeter• has been carefully manufactured
2ContentsSystem description 2Product liability and warranty 2Standards and approvals 2Items included with supply 31 Important information for instal
3Items included with supplyIFS 6000 F primary head• primary headin the size as ordered• certificate of calibration data• installation instructionsIFM
41 Important information for installation: PLEASE NOTE• Location and position as required,but electrode axis X – • – • – • – Xmust be approximately ho
5• Strong electromagnetic fields, avoid in vicinity of flowmeter.• Straight inlet run minimum of 5 ××DN and outlet run minimum of 2 ××DN,(DN = meter s
63 Instrument nameplateTeflon®is a registered trademark of Du PontInstrument nameplate for compact flowmeterssee installation and operating instru
75 Installation in the pipelineLmaxLmin• Installation material, see page 3.• Pipe and operating pressures,see Section 6 “Limits”.• Flange spacing= dim
86 Limits7. Grounding• Warning: All flowmeters must be properly grounded to avoid personnel shock hazard.• The ground conductor should not transm
9Supply power > 50 V AC• Grounding is via the PE protective ground conductor incorporated in the power supplycable, see also Section “Connection to