pH sensorOPTISENS PH 9500Handbook© KROHNE 08/2012 - 4001929601 MA OPTISENS PH 9500 R01 enThe documentation is only complete when used in combination
2 DEVICE DESCRIPTION 10 OPTISENS PH 9500www.krohne.com 08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en The pH sensors are available with different di
INSTALLATION 311OPTISENS PH 9500www.krohne.com08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en3.1 Notes on installation3.2 Storage and transport• S
3 INSTALLATION 12 OPTISENS PH 9500www.krohne.com 08/2012 - 4001929601 MA OPTISENS PH 9500 R01 enStoring the sensor in the provided plastic tube• Scr
INSTALLATION 313OPTISENS PH 9500www.krohne.com08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en3.4 Pre-installation requirementsUnpacking the sensor•
3 INSTALLATION 14 OPTISENS PH 9500www.krohne.com 08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en3.5.1 Connecting the sensor cable to the signal conv
INSTALLATION 315OPTISENS PH 9500www.krohne.com08/2012 - 4001929601 MA OPTISENS PH 9500 R01 enThe following instructions describe the connection of t
3 INSTALLATION 16 OPTISENS PH 9500www.krohne.com 08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en3.5.3 Connecting the cable to the sensorConnecting t
INSTALLATION 317OPTISENS PH 9500www.krohne.com08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en3.7 Installing the sensor 3.7.1 General installation
3 INSTALLATION 18 OPTISENS PH 9500www.krohne.com 08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en3.7.2 Mounting to a flow-through holderInstalling a
INSTALLATION 319OPTISENS PH 9500www.krohne.com08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en3.7.3 Mounting to SENSOFIT IMM 1000 immersion holder3.
All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without the prior written authorisation of KROHNE Messtech
3 INSTALLATION 20 OPTISENS PH 9500www.krohne.com 08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en3.9 Examples of a typical measuring pointThe followi
OPERATION 421OPTISENS PH 9500www.krohne.com08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en4.1 Menu mode structureINFORMATION!The following table ju
4 OPERATION 22 OPTISENS PH 9500www.krohne.com 08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en 3 or 4 pages, scrolling with ↓ or ↑>2.5s^B test >
OPERATION 423OPTISENS PH 9500www.krohne.com08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en4.2 Function tables4.2.1 Menu A, quick setupA9, ph cal.
4 OPERATION 24 OPTISENS PH 9500www.krohne.com 08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en4.2.2 Menu B, testThe procedure to start the simulation
OPERATION 425OPTISENS PH 9500www.krohne.com08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en4.2.3 Menu C, setupC1, process input AC2, process input B
4 OPERATION 26 OPTISENS PH 9500www.krohne.com 08/2012 - 4001929601 MA OPTISENS PH 9500 R01 enC1.15.5C2.15.5temp. comp. Menu for activating the temper
OPERATION 427OPTISENS PH 9500www.krohne.com08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en4.3 Calibration4.3.1 Temperature compensationThere are t
4 OPERATION 28 OPTISENS PH 9500www.krohne.com 08/2012 - 4001929601 MA OPTISENS PH 9500 R01 enAfter starting-up the signal converter, the measuring sc
OPERATION 429OPTISENS PH 9500www.krohne.com08/2012 - 4001929601 MA OPTISENS PH 9500 R01 enStep 2: Configure /adjust the temperature sensor• If nece
CONTENTS 3www.krohne.com08/2012 - 4001929601 MA OPTISENS PH 9500 R01 enOPTISENS PH 95001 Safety instructions 51.1 Intended use ...
4 OPERATION 30 OPTISENS PH 9500www.krohne.com 08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en• Enter the measured temperature. Press ^ to confirm th
OPERATION 431OPTISENS PH 9500www.krohne.com08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en4.3.2 Calibrating pH measurementA pH calibration is neces
4 OPERATION 32 OPTISENS PH 9500www.krohne.com 08/2012 - 4001929601 MA OPTISENS PH 9500 R01 enStep 1: activating the manual hold function• You have a
OPERATION 433OPTISENS PH 9500www.krohne.com08/2012 - 4001929601 MA OPTISENS PH 9500 R01 enStep 3a: accessing the calibration menu via the main menu
4 OPERATION 34 OPTISENS PH 9500www.krohne.com 08/2012 - 4001929601 MA OPTISENS PH 9500 R01 enStep 4: calibration procedure• After choosing the subme
OPERATION 435OPTISENS PH 9500www.krohne.com08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en• Press ^ to confirm.i The slope of the sensor is display
4 OPERATION 36 OPTISENS PH 9500www.krohne.com 08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en4.3.3 Calibration logAccessing the calibtration log• Y
SERVICE 537OPTISENS PH 9500www.krohne.com08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en5.1 Maintenance5.1.1 Cleaning• Slight dirt residues or du
5 SERVICE 38 OPTISENS PH 9500www.krohne.com 08/2012 - 4001929601 MA OPTISENS PH 9500 R01 enThe following figure shows the aging effects:5.2 Spare pa
SERVICE 539OPTISENS PH 9500www.krohne.com08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en5.4 Returning the device to the manufacturer5.4.1 General
CONTENTS 4 www.krohne.com 08/2012 - 4001929601 MA OPTISENS PH 9500 R01 enOPTISENS PH 95005 Service 375.1 Maintenance ...
5 SERVICE 40 OPTISENS PH 9500www.krohne.com 08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en5.4.2 Form (for copying) to accompany a returned device5.
TECHNICAL DATA 641OPTISENS PH 9500www.krohne.com08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en6.1 Measuring principle6.1.1 pH measurementThe meas
6 TECHNICAL DATA 42 OPTISENS PH 9500www.krohne.com 08/2012 - 4001929601 MA OPTISENS PH 9500 R01 enThe voltage change of a pH sensor at 25°C/ 77°F is
TECHNICAL DATA 643OPTISENS PH 9500www.krohne.com08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en6.2 Technical datapH sensorINFORMATION!•The followin
6 TECHNICAL DATA 44 OPTISENS PH 9500www.krohne.com 08/2012 - 4001929601 MA OPTISENS PH 9500 R01 enOperating conditionsTemperature range -5…+100°C / +
TECHNICAL DATA 645OPTISENS PH 9500www.krohne.com08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en6.3 DimensionsFigure 6-4: Dimension of the pH sensor
7 APPENDIX 46 OPTISENS PH 9500www.krohne.com 08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en7.1 pH as a function of mVThe pH-value is the negative d
APPENDIX 747OPTISENS PH 9500www.krohne.com08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en7.2 pH temperature dependencyThe output from a pH-sensor v
KROHNE product overview• Electromagnetic flowmeters• Variable area flowmeters• Ultrasonic flowmeters• Mass flowmeters• Vortex flowmeters• Flow control
SAFETY INSTRUCTIONS 15OPTISENS PH 9500www.krohne.com08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en1.1 Intended useThe intended use of OPTISENS PH
1 SAFETY INSTRUCTIONS 6 OPTISENS PH 9500www.krohne.com 08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en1.2.2 DisclaimerThe manufacturer will not be l
SAFETY INSTRUCTIONS 17OPTISENS PH 9500www.krohne.com08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en1.2.5 Warnings and symbols usedSafety warnings a
2 DEVICE DESCRIPTION 8 OPTISENS PH 9500www.krohne.com 08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en2.1 Scope of deliveryOptional accessories• SENS
DEVICE DESCRIPTION 29OPTISENS PH 9500www.krohne.com08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en2.2 Device description2.2.1 pH sensorFigure 2-2: