Sludge level meterSoftware Revision:01.00.01OPTISYS SLM 2100OPTISYS SLM 2100OPTISYS SLM 2100OPTISYS SLM 2100HandbookHandbookHandbookHandbook© KROHNE 0
2 DEVICE DESCRIPTION 10 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enDevice description2.1 Scope of deliveryINFORM
DEVICE DESCRIPTION 211OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en2.2 Device descriptionMain electronics unitThe
2 DEVICE DESCRIPTION 12 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enCable drum assembly and heater unit The cable
DEVICE DESCRIPTION 213OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en2.3 NameplateINFORMATION!Look at the device nam
3 INSTALLATION 14 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enInstallation3.1 General notes on installation3.2 S
INSTALLATION 315OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en3.3 Installation orderTo install the measuring system
3 INSTALLATION 16 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en3.3.1 Mounting of the sludge level meterThe mountin
INSTALLATION 317OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enStep 2a: Mounting on round handrail• Mount the frame
3 INSTALLATION 18 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enStep 4: Adjusting• Open the door of the meter for 4
INSTALLATION 319OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en3.5 Installing the cleaning unit (optional)The cleani
All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without the prior written authorisation of KROHNE Messtech
3 INSTALLATION 20 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en3.5.1 Installation of the water hoseA 3/4" con
ELECTRICAL CONNECTIONS 421OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enElectrical connections4.1 Safety instructio
4 ELECTRICAL CONNECTIONS 22 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en4.3 Description of electrical symbolsSymb
ELECTRICAL CONNECTIONS 423OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en4.4 Important device-specific notes on elec
4 ELECTRICAL CONNECTIONS 24 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en4.5 Overview cable connectionsIn the bott
ELECTRICAL CONNECTIONS 425OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en4.6 Overview of the terminal compartmentThe
4 ELECTRICAL CONNECTIONS 26 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en4.7 Connecting the power supplyBefore you
ELECTRICAL CONNECTIONS 427OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en4.8 Description and properties of the outpu
4 ELECTRICAL CONNECTIONS 28 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en4.8.2 Control input (active)The electrica
ELECTRICAL CONNECTIONS 429OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enMaintenance modeMaintenance modeMaintenance
CONTENTS 3www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enOPTISYS SLM 21001 Safety instructions 51.1 Software history ...
4 ELECTRICAL CONNECTIONS 30 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en4.9.2 Current output (active)The terminal
ELECTRICAL CONNECTIONS 431OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enConnecting the current output• Open the cabl
4 ELECTRICAL CONNECTIONS 32 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enThe terminals for the connection of the th
ELECTRICAL CONNECTIONS 433OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en4.10 Description and properties of the rela
4 ELECTRICAL CONNECTIONS 34 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enNote the electrical properties of the rela
ELECTRICAL CONNECTIONS 435OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en4.11 Electrical connection of the external
5 OPERATION 36 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enOperation5.1 Start-up and general remarks for configur
OPERATION 537OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en5.2 Switching on the powerSwitch on the powerSwitch on t
5 OPERATION 38 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en5.3 Operating elementsThe operation elements consist o
OPERATION 539OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en5.4 Measuring pageAfter switching on the power, the disp
CONTENTS 4 www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enOPTISYS SLM 21005 Operation 365.1 Start-up and general remarks for conf
5 OPERATION 40 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en5.4.1 Navigating through the menusTo switch from the m
OPERATION 541OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enThe submenu level consists of the first and second submen
5 OPERATION 42 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enOn the parameter level you see a list with one or more
OPERATION 543OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enDisplay with parameter or data ready to editFigure 5-6: E
5 OPERATION 44 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en5.5 Menu mode structureINFORMATION!The following table
OPERATION 545OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en5.6 Function tables5.6.1 Menu A, quick setupA1, languag
5 OPERATION 46 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enB3, logbooksB4, information5.6.3 Menu C, setupC1, meas
OPERATION 547OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enC2, maintenanceC3, I/OLevel Designation / function Settin
5 OPERATION 48 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enC3.4.2 Invert signal This function allows to define ho
OPERATION 549OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enC4, extended setupLevel Designation / function Settings /
SAFETY INSTRUCTIONS 15OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enSafety instructions1.1 Software history1.2 Int
5 OPERATION 50 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enC5, deviceLevel Designation / function Settings / descr
OPERATION 551OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en5.7 Functions in detail5.7.1 Definition of positions an
5 OPERATION 52 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en5.7.2 Measuring functionIn this menu point the measuri
OPERATION 553OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en5.7.4 Definition of fluff and blanket concentrationIn th
5 OPERATION 54 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en5.7.7 Timer / start signalIf there is no danger for th
OPERATION 555OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en5.7.9 CalibrationThere are two forms of calibration. Fir
5 OPERATION 56 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enStep 3: calibrate zeroThe calibration zero is done.• P
OPERATION 557OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enStep 3b: calibrate conc. without known reference valueSte
5 OPERATION 58 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en5.7.10 PasswordsYou can activate passwords for differe
OPERATION 559OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en5.8 Status messages and diagnostic informationThe OPTISY
1 SAFETY INSTRUCTIONS 6 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en1.3 CertificationsThe device fulfils the stat
5 OPERATION 60 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enExplanation of the error categories5.8.1 Error categor
OPERATION 561OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en5.8.2 Error category "Application error" ( &qu
6 SERVICE 62 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enService6.1 Manual initiation of a measurementFor service
SERVICE 663OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enCheck of screws and nutsCheck if all screws of the reeling
6 SERVICE 64 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enCleaning or replacement of air filter1. Loosen the 2 nuts
SERVICE 665OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en6.3 Spare parts availabilityThe manufacturer adheres to th
6 SERVICE 66 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en6.5.2 Form (for copying) to accompany a returned device6
TECHNICAL DATA 767OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enTechnical data7.1 Measuring principleThe meter is u
7 TECHNICAL DATA 68 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en7.2 Technical dataINFORMATION!•The following data
TECHNICAL DATA 769OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enMeasuring accuracyMaximum measuring error (concentra
SAFETY INSTRUCTIONS 17OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en1.4 Safety instructions from the manufacturer1.
7 TECHNICAL DATA 70 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enInputs and outputsGeneral All in and outputs are e
TECHNICAL DATA 771OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enApprovals and certificationsCE The device fulfils th
7 TECHNICAL DATA 72 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en7.3 DimensionsFigure 7-2: DimensionsDimensions [c
NOTES 873OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 enNotes.book Page 73 Wednesday, April 30, 2014 1:39 PM
8 NOTES 74 OPTISYS SLM 2100www.krohne.com 03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en.book Page 74 Wednesday, April 30, 2014 1:39 PM
NOTES 875OPTISYS SLM 2100www.krohne.com03/2014 - 4002737302 - MA OPTISYS SLM 2100 R02 en.book Page 75 Wednesday, April 30, 2014 1:39 PM