7.30830.32.00 Installation andoperatinginstructionsTIDALFLUXIFM 4110 PFElectromagneti

Page 2

7.30830.32.00108 Outer sheathBootstrap signal cable type BTSThe signal converter automatically controls the individual shields (3) to exactly the sam

Page 3

7.30830.32.0011C1203023121110 871 2 3IFC110 PF IFS4000 PFPE N L N L PEInterfaceElectrodesCoilsPowerA B C D E8 778NC NC DNCC121011 ENCNC230/1

Page 4

7.30830.32.00122. Installation of the signal converter2.1 Please note the following information concerning installation and operation of the IFC 110 P

Page 5

7.30830.32.0013• Active mode: the signal converter supplies the power for the operation (selection) of receiver instruments,observe max. operating da

Page 6

7.30830.32.00142.5.3.2 Pulse output A1 for electromechanical totalizersPLEASE NOTE:The output terminal A1 can be used as status output A1 or as a 2nd

Page 7

7.30830.32.0015• Connection diagrams see sect. 2.5.6.• Schematic wiring diagram for status outputs A1/A2 and D1/D2.These status outputs have MOSFET

Page 8

7.30830.32.0016PPD1D⊥D2C1C⊥C2I+I-A1A⊥A2E+E-I IContacts 24 V, 10 mAI ≤ 7 mAControl inputsC1 / C2 activePPD1D⊥D2C1C⊥C2I+I-A1A⊥A2E+E-I I _∼∼UextUext ≤ 32

Page 9

7.30830.32.0017reverseflowforwardflowElectronic totalizers must be connected as shown in the connection diagrams for pulse output Pon the following fi

Page 10

7.30830.32.0018* Shielded cables must be used to prevent radio interference at pulse output frequencies > 100 Hz.for frequencies ≤ 1 kHzUext. ≤ 3

Page 11 - IFS4000 PF

7.30830.32.00192.5.7 Standard factory settingsAll operating data are set at the factory in accordance with the specifications contained in the order.

Page 12

7.30830.32.002ContentsProduct liability and warrantySystem descriptionAvailable versionItems included with supplyCE / EMC / Standards / Certificationp

Page 13

7.30830.32.00203 Start-up• Before connecting to power, check that the instrument is correctly installed as described in Sections 1 and 2.• The flow

Page 14

7.30830.32.0021Part B Signal converter IFC 110 PF4. Operating the signal converter4.1 Krohne operating conceptMeasuring modeWhen this is shown on the

Page 15

7.30830.32.00224.2 Operating and control elements…the 15 keys à and Ä accessibleafter removal of the glass cover,…the 3 magnetic sensors Ç and the

Page 16 - ≥ 160 Ω I ≤ 100 mA

7.30830.32.00234.3 Key functionsIn the following, the cursor or flashing part of the display is shown against a gray background.To start operator con

Page 17 - 7.30830.32.0017

7.30830.32.0024Shift to leftTo alter text (units)In case of units, the numericalvalue is converted automatically.Select next textSelect preceding text

Page 18

7.30830.32.0025Fct. No. Text Description and setting1.0 OPERATION Operating menuFULL SCALE Full-scale range for flow rate Q100%Selection of unit• m3/h

Page 19

7.30830.32.0026Press ↵ key to change to subfunction “RANGE I”→ RANGE I Selecting the measuring range• 0-20 mA • 4-20 mA (fixed ranges)• mA (user-defin

Page 20

7.30830.32.0027→ • OFF • ON • ALL ERROR • FATAL.ERROR• INVERS D1 (inverse mode of D1 and D2)• INVERS A1 (inverse mode of A1 or A2 possible only if A1

Page 21 - Direction of

7.30830.32.0028Fct. Text Description and settings3.0 INSTALL. Installation menu3.01 LANGUAGE Select language for display texts• GB / USA (English) • F

Page 22 - 6 6 6 6 6 6

7.30830.32.0029→ ADC GAIN Set gain of A/D converter• AUTO • 10 • 30 • 100 Select with ↑ or ↓ keyPress ↵ key to change to subfunction “SPEC. FILT.”.→ S

Page 23 - Press key

7.30830.32.0035.3 Low-flow cutoff SMU5.4 Display5.5 Internal electronic counter5.6 Internal power supply (E+ / E-) for connected loads5.7 Curren

Page 24

7.30830.32.0030ADC-HARDW. Hardware error A/D converter Replace ADC printed circuit boardADC GAIN Hardware error A/D converter Replace ADC printed circ

Page 25 - 7.30830.32.0025

7.30830.32.00315. Description of functions5.1 Full-scale range Q100%Fct. 1.01 FULL SCALEPress → keySelect unit for full-scale range Q100%•• m3/hr (cub

Page 26 - 7.30830.32.0026

7.30830.32.00325.3 Low-flow cutoff SMUFct. 1.03 L.F. CUTOFFPress → keySelect• OFF (fixed tripping points: ON = 0.1 % / OFF = 0.2 %)• PERCENT (variabl

Page 27 - 7.30830.32.0027

7.30830.32.0033Setting of totalizer format• Auto (exponent notation)• # . # # # # # # # • # # # # # . # # #• # # . # # # # # # • # # # # # # . # #• #

Page 28 - 7.30830.32.0028

7.30830.32.00345.6 Internal power supply (E+ / E-) for connected loadsPassive loads connected to the outputs and inputs can be fed by means of the int

Page 29

7.30830.32.0035→→ I ERROR = set error value, press → key• 22 mA fixed value• 0.0 – I0% mA variable value, only variable when I0% ≥ 1 mA, see “RANGE I”

Page 30

7.30830.32.0036→→ PULSWIDTH = select pulse width, press → key• AUTO automatic = 50 % of cycle duration of 100 % output frequency• SYM. symmetric = pu

Page 31

7.30830.32.00375.9 Status outputs A1 / A2 and D1 / D2PLEASE NOTE:Connection diagrams see sect. 2.5.6.Status outputs A1 A2 D1 D2Select Fct. _ . _ _the

Page 32

7.30830.32.0038•• Characteristic of status outputs Switch open Switch closedOFF (switched off)no functionON (e.g. operation indicator)power supply OFF

Page 33

7.30830.32.0039Select with ↑ and ↓ keys.Press ↵ key to return to Fct. 3.01 LANGUAGE.5.12 Entry codeFct. 3.04 ENTRY CODEPress → keySelect• NO no code,

Page 34

7.30830.32.00410.2.1 General information10.2.2 Dimensions and weight IFC 110 PF10.3 Complete system IFM 4110 PF10.3.1 Full-scale range Q100%10.3.2

Page 35

7.30830.32.0040→→ LINE FREQ. = set power frequency customary in country where instrument is used, press → key(Please note: only applies to instruments

Page 36

7.30830.32.0041Factors for volume FMVolumetric unit TEXT VOL. Factor FMSettingCubic metersm31.01.00000 E+0LiterLiter10001.00000 E+3Hectolitersh Lit101

Page 37

7.30830.32.00425.16 Output characteristicsI current outputI0%0 or 4 mAI100%20 mAP pulse outputs P and A1 (P2)P100%pulses at Q100%, full-scale rangeQF

Page 38

7.30830.32.00435.17 ApplicationsFct. 3.06 APPLICAT.Press → key twiceSet flow characteristics, select with ↑ or ↓ keys• STEADY steady flow• PULSATING

Page 39

7.30830.32.00445.19 Limit switchesFct. 1.07 – 1.10 Status outputs A1, A2, D1 and D2(Define operating mode of output terminals A1, see sect. 5.18)Pres

Page 40

7.30830.32.0045Press ↵ key to return to Fct. 1.07 – 1.10, status outputs A1, A2, D1 or D2.External range change by control inputFct. 1.11 or 1.12 Cont

Page 41

7.30830.32.0046Part C Special Applications, Functional Checks,Service and Order Numbers6. Special applications6.1 Use in hazardous areasThe IFS 4000 P

Page 42

7.30830.32.0047Resetting the signal converter, see sections 4 and 5.When changing the signal converter settings, the green LED (normal) and the red LE

Page 43

7.30830.32.0048Set subfunction “LIMIT CNT.” to 10 on a trial basis.Should the measuring value leave the above window in one direction more than 10 ti

Page 44 - XXX YYY

7.30830.32.00497. Functional checks7.1 Checking the zero with IFC 110 PF signal converter, Fct. 3.03• Set “zero” flow in the pipeline. Make sure that

Page 45

7.30830.32.005How to use these Installation and Operating Instructions• For easy reference these instructions are divided into 5parts.• Only Part A

Page 46 - Service and Order Numbers

7.30830.32.0050 allow your flowmeter to be subjected to a simple and rapid diagnosis.• Press the following keys for a display of the status codes:Ke

Page 47

7.30830.32.00517.4 Hardware test, Fct. 2.03Please note:Before beginning the test, deactivate any alarms and controllers as the current output will be

Page 48

7.30830.32.0052LED 2Red LED flashes Fatal error, hardware and/orsoftware faultReplace signal converter, seesect. 8.3LED 3Cyclic flashing of red LED,ap

Page 49

7.30830.32.0053Group I Faults / Symptoms Cause Remedial actionDisplay shows…I SHORTCurrent output shorted,load < 15 ΩEliminate short circuit,load m

Page 50

7.30830.32.0054Group P Faults / Symptoms Cause Remedial actionWrong connection / polarity Connect properly, see sect.2.5.3 and 2.5.6, noterecommended

Page 51

7.30830.32.0055Group S Faults / Symptoms Cause Remedial actionDefective signaling instrument(s)or external voltage sourceCheck signaling instrument(s)

Page 52

7.30830.32.00567.6 Checking the primary headSince the primary head consists of two “separated” parts (velocity measurement and level measurement), the

Page 53

7.30830.32.0057electrodes contaminated.4When signal line BTS (bootstrap) isused:measure resistance between thefollowing lines:1 and 20 / 1 and 30 / 20

Page 54

7.30830.32.0058Switch off the power supply before starting work.1) Remove the cover from the terminal compartment of the signal converter.2) Disconn

Page 55

7.30830.32.00594.2) Determine position of switch Y: use table (GS 8 A front panel) to determine value Y which comes closest to factor X andmeets condi

Page 56

7.30830.32.006Available versionSystem: IFM 4110 PFPrimary head:• type: IFS 4000 PF• liner measuring section: Irathane• Meter size: 200 - 600 mm /

Page 57

7.30830.32.00604.3) Calculate setpoint reading “I” for current output: I = I0% + YX (I100% - I0%) in mAI0%current (0/4 mA) at 0% flow rateI100%current

Page 58

7.30830.32.0061If you need to return your flowmeter to Krohne, please refer to the last-but-one page of these instructions.

Page 59 - 7.30830.32.0059

7.30830.32.00628. Service8.1 Replacing the power supply fusePower supply fuse in IFC 110 PF signal converterSwitch off the power supply before openi

Page 60

7.30830.32.00638.3 Replacing the complete electronic unit of the IFC 110 PF signal converterSwitch off the power supply before opening the housing.1)

Page 61 - 7.30830.32.0061

7.30830.32.00648.4 Replacing single printed circuit boards (PCBs)Switch off the power supply before opening the housing.1) Remove the cover from the

Page 62

7.30830.32.0065Jumper X4DC operation ≤ 0.2 AAC operation ≤ 0.1 A(factory setting)X1 plug-in terminals inside terminal compartmentX2 internal connectio

Page 63

7.30830.32.0066Analog/digital converter PCB, ADCX1 internal connection to motherboard IC 12 microprocessorX2 plug-in terminals in terminal compartment

Page 64

7.30830.32.00679 Order numbersSpare parts Order No.Electronic unit with display 100-230 V AC without magnetic sensors 2106680000100-230 V AC with magn

Page 65 - + +

7.30830.32.0068Part D Technical Data, Measuring Principle and Block Diagram10. Technical data10.1 Primary head IFS 4000 PF10.1.1 General informationM

Page 66

7.30830.32.006910.1.2 Dimensions and weight IFS 4000 PFMeter size to … Dimensions in mm (inches)DIN 2501Approx.weightMmPNANSIB16.5a b ∅c D j ∅D ∅Di kg

Page 67

7.30830.32.007Part A System installation and start-up1 Installation of the primary head1.1 Selecting the installation location1. Location and positi

Page 68

7.30830.32.0070Pulse outputs (passive) P A1(also used as status output)- for electronic totalizers - for electromechanical totalizers- all operating d

Page 69

7.30830.32.0071Local display 3-line LCDDisplay function current flow, forward, reverse and sum totalizers (7-digit) or 25-digit bar graph withpercenta

Page 70

7.30830.32.007210.2.2 Dimension and weight IFC 110 PFWeight approx. 4.1 kg (9.0 lb)Dimensions in mm (inch)10.3 Complete system IFM 4110 PF10.3.1 Full

Page 71

7.30830.32.0073Flow tablev = flow velocity in m/s v = flow velocity in ft/sMeter size Full-scale range in m3/hr Meter size Full-scale range in US Gal/

Page 72

7.30830.32.007411 Block diagramAD µµPEEPROM 1 EEPROM 2EEPROM-EEPROM µµP• LC display• 15 keys• 3 magnetic sensors (optional)KSA 04 µµPEEPROMP

Page 73

7.30830.32.0075ÀÀ ADC printed circuit board, analog/digital converter (terminals 1, 2, 3, 20 and 30)• Signal processor protected against overload, f

Page 74 - EEPROM 1 EEPROM 2

7.30830.32.007612 Measuring principleThe TIDALFLUX IFM 4110 PF is an electromagnetic flowmeter with an integrated capacitive level measurement system,

Page 75

7.30830.32.0077Part E IndexKeyword Sect. No. Fct. No. Keyword Sect. No. Fct. No.AA1 status or 2nd pulse outputA1/A2 status outputsAbbreviationsADC see

Page 76

7.30830.32.0078Electrical connection- GS 8 A simulator- inputs- outputs- power supply- primary headElectromagnetic compatibility- setting- contro

Page 77 - Part E Index

7.30830.32.0079PP = pulse outputsPCB = printed circuit boardPC softwarePE = protective conductorPower supply- connection- failure- frequency- inpu

Page 78

7.30830.32.008350 10 16x M20 75 (54.2)400 10 16x M24 104 (75.2)500 10 20x M24 107 (77.4)600 10 20x M27 138 (99.8)MetersizeinchesBodypressureratingpsig

Page 79

7.30830.32.0080If you need to return flowmeters for testing or repairing to KrohneYour electromagnetic flowmeter has been carefullymanufactured and te

Page 80

7.30830.32.0091.5 Electrical connections of primary head1.5.1 Connection to powerElectrical connection in conformity with VDE 0100 / EN 61010-1“Regu