Krohne OPTISENS PH 8590 User Manual

Browse online or download User Manual for Equipment Krohne OPTISENS PH 8590. KROHNE OPTISENS PH 8590 User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 46
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
pH sensor
OPTISENS PH 8590
OPTISENS PH 8590OPTISENS PH 8590
OPTISENS PH 8590
Handbook
HandbookHandbook
Handbook
© KROHNE 03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en
.book Page 1 Monday, March 19, 2012 2:01 PM
Page view 0
1 2 3 4 5 6 ... 45 46

Summary of Contents

Page 1 - OPTISENS PH 8590

pH sensorOPTISENS PH 8590OPTISENS PH 8590OPTISENS PH 8590OPTISENS PH 8590HandbookHandbookHandbookHandbook© KROHNE 03/2012 - 4001931401 MA OPTISENS PH

Page 2

2 DEVICE DESCRIPTION 10 OPTISENS PH 8590www.krohne.com 03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en The pH electrodes are available with different

Page 3 - CONTENTS

INSTALLATION 311OPTISENS PH 8590www.krohne.com03/2012 - 4001931401 MA OPTISENS PH 8590 R01 enInstallation3.1 Notes on installation3.2 Storage and t

Page 4

3 INSTALLATION 12 OPTISENS PH 8590www.krohne.com 03/2012 - 4001931401 MA OPTISENS PH 8590 R01 enStoring the electrode in the provided plastic tube• T

Page 5 - SAFETY INSTRUCTIONS 1

INSTALLATION 313OPTISENS PH 8590www.krohne.com03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en3.4 Pre-installation requirementsUnpacking the electrod

Page 6 - SAFETY INSTRUCTIONS

3 INSTALLATION 14 OPTISENS PH 8590www.krohne.com 03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en3.5.1 Connecting the sensor cable to the signal conve

Page 7

INSTALLATION 315OPTISENS PH 8590www.krohne.com03/2012 - 4001931401 MA OPTISENS PH 8590 R01 enThe following instructions describe the connection of th

Page 8

3 INSTALLATION 16 OPTISENS PH 8590www.krohne.com 03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en3.5.3 Connecting the cable to the sensorConnecting th

Page 9 - DEVICE DESCRIPTION 2

INSTALLATION 317OPTISENS PH 8590www.krohne.com03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en3.7 Installing the sensor 3.7.1 General installation i

Page 10 - DEVICE DESCRIPTION

3 INSTALLATION 18 OPTISENS PH 8590www.krohne.com 03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en3.8 Installing an external temperature sensorSince th

Page 11 - INSTALLATION 3

OPERATION 419OPTISENS PH 8590www.krohne.com03/2012 - 4001931401 MA OPTISENS PH 8590 R01 enOperation4.1 Menu mode structureINFORMATION!The following

Page 12 - INSTALLATION

All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without the prior written authorisation of KROHNE Messtech

Page 13

4 OPERATION 20 OPTISENS PH 8590www.krohne.com 03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en 3 or 4 pages, scrolling with ↓ or ↑>2.5s^B testB test

Page 14

OPERATION 421OPTISENS PH 8590www.krohne.com03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en4.2 Function tables4.2.1 Menu A, quick setupA9, ph cal. B

Page 15

4 OPERATION 22 OPTISENS PH 8590www.krohne.com 03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en4.2.2 Menu B, testThe procedure to start the simulation

Page 16

OPERATION 423OPTISENS PH 8590www.krohne.com03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en4.2.3 Menu C, setupC1, process input AC1, process input AC

Page 17

4 OPERATION 24 OPTISENS PH 8590www.krohne.com 03/2012 - 4001931401 MA OPTISENS PH 8590 R01 enC1.15.5C2.15.5temp. comp. Menu for activating the tempera

Page 18

OPERATION 425OPTISENS PH 8590www.krohne.com03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en4.3 Calibration4.3.1 Temperature compensationThere are th

Page 19 - OPERATION 4

4 OPERATION 26 OPTISENS PH 8590www.krohne.com 03/2012 - 4001931401 MA OPTISENS PH 8590 R01 enAfter starting-up the signal converter, the measuring scr

Page 20 - OPERATION

OPERATION 427OPTISENS PH 8590www.krohne.com03/2012 - 4001931401 MA OPTISENS PH 8590 R01 enStep 2: Configure /adjust the temperature sensor• If neces

Page 21

4 OPERATION 28 OPTISENS PH 8590www.krohne.com 03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en• Enter the measured temperature. Press ^ to confirm the

Page 22

OPERATION 429OPTISENS PH 8590www.krohne.com03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en4.3.2 Calibrating pH measurementA pH calibration is necess

Page 23

CONTENTS 3www.krohne.com03/2012 - 4001931401 MA OPTISENS PH 8590 R01 enOPTISENS PH 85901 Safety instructions 51.1 Intended use ...

Page 24

4 OPERATION 30 OPTISENS PH 8590www.krohne.com 03/2012 - 4001931401 MA OPTISENS PH 8590 R01 enStep 1: activating the manual hold function• You have ac

Page 25

OPERATION 431OPTISENS PH 8590www.krohne.com03/2012 - 4001931401 MA OPTISENS PH 8590 R01 enStep 3a: accessing the calibration menu via the main menu s

Page 26

4 OPERATION 32 OPTISENS PH 8590www.krohne.com 03/2012 - 4001931401 MA OPTISENS PH 8590 R01 enStep 4: calibration procedure• After choosing the submen

Page 27

OPERATION 433OPTISENS PH 8590www.krohne.com03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en• Press ^^^^ to confirm.i The slope of the electrode is di

Page 28

4 OPERATION 34 OPTISENS PH 8590www.krohne.com 03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en4.3.3 Calibration logAccessing the calibtration log• Yo

Page 29

SERVICE 535OPTISENS PH 8590www.krohne.com03/2012 - 4001931401 MA OPTISENS PH 8590 R01 enService5.1 Maintenance5.1.1 Cleaning• Slight dirt residues

Page 30

5 SERVICE 36 OPTISENS PH 8590www.krohne.com 03/2012 - 4001931401 MA OPTISENS PH 8590 R01 enThe following figure shows the aging effects:5.2 Spare par

Page 31

SERVICE 537OPTISENS PH 8590www.krohne.com03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en5.4 Returning the device to the manufacturer5.4.1 General i

Page 32

5 SERVICE 38 OPTISENS PH 8590www.krohne.com 03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en5.4.2 Form (for copying) to accompany a returned device5.5

Page 33

TECHNICAL DATA 639OPTISENS PH 8590www.krohne.com03/2012 - 4001931401 MA OPTISENS PH 8590 R01 enTechnical data6.1 Measuring principle6.1.1 pH measur

Page 34

CONTENTS 4 www.krohne.com 03/2012 - 4001931401 MA OPTISENS PH 8590 R01 enOPTISENS PH 85905 Service 355.1 Maintenance ...

Page 35 - SERVICE 5

6 TECHNICAL DATA 40 OPTISENS PH 8590www.krohne.com 03/2012 - 4001931401 MA OPTISENS PH 8590 R01 enThe voltage change of a pH electrode at 25°C/ 77°F i

Page 36 - SERVICE

TECHNICAL DATA 641OPTISENS PH 8590www.krohne.com03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en6.2 Technical datapH electrodesINFORMATION!•The follo

Page 37

6 TECHNICAL DATA 42 OPTISENS PH 8590www.krohne.com 03/2012 - 4001931401 MA OPTISENS PH 8590 R01 enOperating conditionsTemperature range 0…+80°C / +23…

Page 38

TECHNICAL DATA 643OPTISENS PH 8590www.krohne.com03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en6.3 DimensionsFigure 6-4: Dimensions OPTISENS PH 8590

Page 39 - TECHNICAL DATA 6

7 APPENDIX 44 OPTISENS PH 8590www.krohne.com 03/2012 - 4001931401 MA OPTISENS PH 8590 R01 enAppendix7.1 pH as a function of mVThe pH-value is the neg

Page 40 - TECHNICAL DATA

APPENDIX 745OPTISENS PH 8590www.krohne.com03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en7.2 pH temperature dependencyThe output from a pH-electrode

Page 41

KROHNE product overview• Electromagnetic flowmeters• Variable area flowmeters• Ultrasonic flowmeters• Mass flowmeters• Vortex flowmeters• Flow control

Page 42

SAFETY INSTRUCTIONS 15OPTISENS PH 8590www.krohne.com03/2012 - 4001931401 MA OPTISENS PH 8590 R01 enSafety instructions1.1 Intended useThe intended u

Page 43

1 SAFETY INSTRUCTIONS 6 OPTISENS PH 8590www.krohne.com 03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en1.2.2 DisclaimerThe manufacturer will not be li

Page 44 - APPENDIX

SAFETY INSTRUCTIONS 17OPTISENS PH 8590www.krohne.com03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en1.2.5 Warnings and symbols usedSafety warnings ar

Page 45 - APPENDIX 7

2 DEVICE DESCRIPTION 8 OPTISENS PH 8590www.krohne.com 03/2012 - 4001931401 MA OPTISENS PH 8590 R01 enDevice description2.1 Scope of deliveryOptional

Page 46 - KROHNE product overview

DEVICE DESCRIPTION 29OPTISENS PH 8590www.krohne.com03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en2.2 Device description2.2.1 pH electrodesThe OPTI