Krohne OPTISOUND 3010C 4wire EN User Manual

Browse online or download User Manual for Equipment Krohne OPTISOUND 3010C 4wire EN. KROHNE OPTISOUND 3010C 4wire EN User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - Operating Instructions

Operating InstructionsOPTISOUND 3010 C4 ... 20 mA/HART - four-wire

Page 2 - Contents

4 Mounting4.1 General instructionsSelect an installation position you can easily reach formounting and connecting as well as later retrofitting of anin

Page 3

12Fig. 3: Min. distance to the max. level1 Dead zone2 Reference planeInformation:If the product reaches the transducer, buildup can form on itover a p

Page 4 - 1 Sequence

4.2 Mounting informationScrew OPTISOUND 3010 C into the mounting socket with anappropriate spanner applied to the hexagon of the processfitting. Max. t

Page 5 - 2 For your safety

Fig. 6: Vessel with conical bottomThe mounting socket should preferably be dimensioned toallow the lower edge of the transducer to protrude at least 1

Page 6

> 10 mmdhdh100 mm/4"150 mm/6"300 mm/12"400 mm/16"Fig. 8: Deviating socket dimensionsIn liquids, direct the sensor as perpendicu

Page 7 - 3 Product description

The ultrasonic sensor should be installed at a location whereno installations cross the ultrasonic beam.Vessel installations such as for example, ladd

Page 8 - 3.4 Storage and transport

Fig. 12: AgitatorsDo not mount the instruments in or above the filling stream.Make sure that you detect the product surface and not theinflowing product

Page 9 - Relative humidity 20 … 85 %

Guided radar is not influenced by foam generation and isparticularly suitable for such applications.If there are strong air currents in the vessel, e.g

Page 10 - 4 Mounting

5 Connecting to power supply5.1 Preparing the connectionAlways observe the following safety instructions:l Connect only in the complete absence of lin

Page 11 - 30504-EN-050914

If potential equalisation currents are expected, the connectionon the evaluation side must be made via a ceramic capacitor(e.g. 1 nF, 1500 V). The low

Page 12 - 4.2 Mounting information

Contents1 About this document1.1 Function ...51.2 Target group ...51.3 Symbolism used ...

Page 13

Fig. 15: Connection steps 5 and 65.3 Wiring plans, double chamber housing1 2 345Fig. 16: Double chamber housing1 Housing cover, connection compartment

Page 14

1Display12 5678I2CFig. 17: Electronics compartment, double chamber housing1 Internal connection cable to the connection compartment12Display12I2CFig.

Page 15

4 ... 20 mAPE/ L/ NL1 NGND1 212344...20mA ISFig. 19: Wiring plan, double chamber housing1 Supply voltage2 Signal output5.4 Switch on phaseAfter OPTISO

Page 16

6 Setup with the indicating/adjustmentmodule6.1 Short descriptionThe indicating/adjustment module is used for measured valuedisplay, adjustment and di

Page 17

Fig. 20: Installation of the indicating/adjustment moduleNote:If you intend to retrofit OPTISOUND 3010 C with an indicating/adjustment module for conti

Page 18 - 5 Connecting to power supply

6.3 Adjustment system1.1231Fig. 21: Indicating and adjustment elements1 LC display2 Indication of the menu item number3 Adjustment keysl [OK] key:- mo

Page 19 - 5.2 Connection procedure

6.4 Setup procedureIn HART-Multidrop mode (several sensors on one input) theaddress must be set before continuing with the parameteradjustment. You wi

Page 20

▶Basic adjustmentDisplayDiagnosticsServiceInfo2 Select the menu item "Basic adjustment" with [–>] andconfirm with [OK]. Now the menu item

Page 21

vessel wall (with solids). To adapt the sensor to these differentconditions, you should first select in this menu item "Liquid" or"Solid&

Page 22 - 5.4 Switch on phase

Enter the requested parameter via the appropriate keys, saveyour settings and jump to the next menu item with the [–>] key.A linearization is neces

Page 23 - 6.1 Short description

8 Dismounting8.1 Dismounting procedure ...378.2 Disposal ...379 Supplement9.1 Technical data...

Page 24

False signal suppressionChange now?Proceed as follows:1 Move from the measured value display to the main menuby pushing [OK].2 Select the menu item &q

Page 25 - - modify value of a parameter

6.5 Menu schematicBasic adjustment1▶Basic adjustmentDisplayDiagnosticsServiceInfo1.1Min. adjustment0.00 %=4.000 m(d)3.000 m(d)1.2Max. adjustment100.00

Page 26 - Address 0

Service4Basic adjustmentDisplayDiagnostics▶ServiceInfo4.1False signal suppressionChange now?4.2Addl. adjustmentsnone4.3Current outputOutput mode 4-20

Page 27

7 Maintenance and fault rectification7.1 MaintenanceWhen used as directed in normal operation, OPTISOUND3010 C is completely maintenance-free.7.2 Fault

Page 28

? Current signal greater than 22 mA or less than 3.6 mAl Electronics module defectiveà Exchange instrument or return it for repairIn Ex applications,

Page 29 - Sensor-TAG

If there is no electronics module available on site, it can beordered from the responsible Krohne agency.7.4 Instrument repairIf a repair is necessary

Page 30 - Change now?

8 Dismounting8.1 Dismounting procedureWarning:Before dismounting, be aware of dangerous process con-ditions such as e.g. pressure in the vessel, high

Page 31 - 6.5 Menu schematic

9 Supplement9.1 Technical dataGeneral dataMaterials, wetted parts- Process fitting Thread G1½A and 1½ NPT: PVDF- Transducer PVDF- Seal transducer/proce

Page 32

Accuracy (similar to DIN EN 60770-1)Reference conditions acc. to DIN EN 61298-1- Temperature 18 … +30°C (64 … +86°F)- Relative humidity 45 … 75 %- Atm

Page 33 - 7.2 Fault rectification

Process conditionsVessel pressure -20 … 200 kPa (-0.2 … 2.0 bar/-2.9 … 29 psi)Process temperature (transducer tem-perature)-40 … +80°C (-40 … +176°F)V

Page 34

1 About this document1.1 FunctionThis operating instructions manual has all the information youneed for quick setup and safe operation of OPTISOUND 30

Page 35 - 7.4 Instrument repair

Approvals6)7)ATEX ATEX II 1G, 1/2G, 2GEExiaIIC T66)Deviating data with Ex applications: see separate safety instructions.7)Depending on order specifica

Page 36 - 8 Dismounting

9.2 DimensionsOPTISOUND 3010 C1 2155mm (6 7/64")22mm (55/64")66mm(2 19/32")ø 39mm(1 17/32")ø 74mm(2 58/64")60mm(2 23/64"

Page 37 - 9 Supplement

9.3 CertificateCE declaration of conformityFig. 24: CE declaration of conformity42 OPTISOUND 3010 C - 4 ... 20 mA/HART - four-wireSupplement30504-EN-05

Page 38

OPTISOUND 3010 C - 4 ... 20 mA/HART - four-wire 43Supplement30504-EN-050914

Page 39

Subject to change without notice30504-EN-050914

Page 40 - Approvals

2 For your safety2.1 Authorised personnelAll operations described in this operating instructions manualmust be carried out only by trained, specialise

Page 41 - 9.2 Dimensions

The parameter adjustment of the basic sensor functions isindependent of the software version. The range of availablefunctions depends on the respectiv

Page 42 - 9.3 Certificate

3 Product description3.1 ConfigurationThe scope of delivery encompasses:l OPTISOUND 3010 C ultrasonic sensorl Documentation- this operating instruction

Page 43

3.2 Principle of operationOPTISOUND 3010 C is an ultrasonic sensor for continuouslevel measurement. It is suitable for liquids and solids invirtually

Page 44

l Relative humidity 20 … 85 %OPTISOUND 3010 C - 4 ... 20 mA/HART - four-wire 9Product description30504-EN-050914