Free chlorine/chlorine dioxide/ozone measuring system
OPTISYS CL 1100
Handbook
© KROHNE 01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en
The documentation is only complete when used in combination with the relevant
documentation for the signal converter.
Page view 0
1 2 3 4 5 6 ... 43 44

Summary of Contents

Page 1 - OPTISYS CL 1100

Free chlorine/chlorine dioxide/ozone measuring systemOPTISYS CL 1100Handbook© KROHNE 01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 enThe documentation

Page 2 - Copyright 2012 by

2 DEVICE DESCRIPTION 10 OPTISYS CL 1100www.krohne.com 01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en2.3 Device description of the sensorThe sensor

Page 3 - CONTENTS

INSTALLATION 311OPTISYS CL 1100www.krohne.com01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en3.1 Notes on installation3.2 Pre-installation requirem

Page 4

3 INSTALLATION 12 OPTISYS CL 1100www.krohne.com 01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en3.3 Installing the sensor 3.3.1 Mounting to OPTISYS

Page 5 - SAFETY INSTRUCTIONS 1

ELECTRICAL CONNECTIONS 413OPTISYS CL 1100www.krohne.com01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en4.1 Safety instructions4.2 Connecting the ca

Page 6 - SAFETY INSTRUCTIONS

4 ELECTRICAL CONNECTIONS 14 OPTISYS CL 1100www.krohne.com 01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en4.3 Connecting the sensor cable to the sign

Page 7

ELECTRICAL CONNECTIONS 415OPTISYS CL 1100www.krohne.com01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 enConnecting the sensor cable to the signal conv

Page 8

5 OPERATION 16 OPTISYS CL 1100www.krohne.com 01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en5.1 Menu mode structureINFORMATION!The following table j

Page 9 - DEVICE DESCRIPTION 2

OPERATION 517OPTISYS CL 1100www.krohne.com01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 enB4 actual values B4.1 operating hoursB4.2 process input AB4

Page 10 - DEVICE DESCRIPTION

5 OPERATION 18 OPTISYS CL 1100www.krohne.com 01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en5.2 Function tables5.2.1 Menu A, quick setupA9, process

Page 11 - INSTALLATION 3

OPERATION 519OPTISYS CL 1100www.krohne.com01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en5.2.2 Menu B, testThe procedure to start the simulation pr

Page 12 - INSTALLATION

All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without the prior written authorisation of KROHNE Messtech

Page 13 - ELECTRICAL CONNECTIONS 4

5 OPERATION 20 OPTISYS CL 1100www.krohne.com 01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en5.2.3 Menu C, setupC1, process input AC2, process input

Page 14 - ELECTRICAL CONNECTIONS

OPERATION 521OPTISYS CL 1100www.krohne.com01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 enC1.13.4 clean Start cleaning procedure, manually.C1.15C2.15

Page 15

5 OPERATION 22 OPTISYS CL 1100www.krohne.com 01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en5.3 Calibration5.3.1 Calibration orderIf you are also m

Page 16 - OPERATION

OPERATION 523OPTISYS CL 1100www.krohne.com01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 enAfter starting-up the signal converter, the measuring scree

Page 17 - OPERATION 5

5 OPERATION 24 OPTISYS CL 1100www.krohne.com 01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 enPress ^ to confirm the entered value. The temperature sen

Page 18

OPERATION 525OPTISYS CL 1100www.krohne.com01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 enEnter the measured temperature.Press ^ to confirm the enter

Page 19

5 OPERATION 26 OPTISYS CL 1100www.krohne.com 01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 enPlease consider that you can activate the pH compensation

Page 20

OPERATION 527OPTISYS CL 1100www.krohne.com01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 enStep 2: entering a pH value manually• If you have activate

Page 21

5 OPERATION 28 OPTISYS CL 1100www.krohne.com 01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 enStep 1: activating the manual hold functionYou have activ

Page 22

OPERATION 529OPTISYS CL 1100www.krohne.com01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 enAfter activating the manual hold function and the preparati

Page 23

CONTENTS 3www.krohne.com01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 enOPTISYS CL 11001 Safety instructions 51.1 Intended use ...

Page 24

5 OPERATION 30 OPTISYS CL 1100www.krohne.com 01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 enStep 4: calibration procedure• After choosing the submen

Page 25

OPERATION 531OPTISYS CL 1100www.krohne.com01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 enCompleting the calibration procedure• As soon as the labor

Page 26

5 OPERATION 32 OPTISYS CL 1100www.krohne.com 01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en5.3.5 Calibration logAccessing the calibtration log• Yo

Page 27

SERVICE 633OPTISYS CL 1100www.krohne.com01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en6.1 Maintenance6.1.1 CleaningThe sensor can be equipped wit

Page 28

6 SERVICE 34 OPTISYS CL 1100www.krohne.com 01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en6.1.2 Aging and re-calibrationThe sensor has to be re-cali

Page 29

SERVICE 635OPTISYS CL 1100www.krohne.com01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en6.4 Returning the device to the manufacturer6.4.1 General i

Page 30

6 SERVICE 36 OPTISYS CL 1100www.krohne.com 01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en6.4.2 Form (for copying) to accompany a returned device6.5

Page 31

TECHNICAL DATA 737OPTISYS CL 1100www.krohne.com01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en7.1 Measuring principleThe sensor has three electrode

Page 32

7 TECHNICAL DATA 38 OPTISYS CL 1100www.krohne.com 01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 enFree chlorine (chlorine dissolved in water) changes

Page 33 - SERVICE 6

TECHNICAL DATA 739OPTISYS CL 1100www.krohne.com01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 enChlorine dioxide measurementChlorine dioxide (ClO2) is

Page 34 - SERVICE

CONTENTS 4 www.krohne.com 01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 enOPTISYS CL 11006 Service 336.1 Maintenance ...

Page 35

7 TECHNICAL DATA 40 OPTISYS CL 1100www.krohne.com 01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 enOzone measurementOzone (O3) is an instable molecule

Page 36

TECHNICAL DATA 741OPTISYS CL 1100www.krohne.com01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en7.2 Technical dataTechnical data for the systemOPTISY

Page 37 - TECHNICAL DATA 7

7 TECHNICAL DATA 42 OPTISYS CL 1100www.krohne.com 01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 enTechnical data for the sensorFor further information

Page 38 - TECHNICAL DATA

TECHNICAL DATA 743OPTISYS CL 1100www.krohne.com01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en7.3 Dimensions and weights for the systemWeight: 5.15

Page 39

KROHNE product overview• Electromagnetic flowmeters• Variable area flowmeters• Ultrasonic flowmeters• Mass flowmeters• Vortex flowmeters• Flow control

Page 40

SAFETY INSTRUCTIONS 15OPTISYS CL 1100www.krohne.com01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en1.1 Intended useThe intended use of the OPTISYS C

Page 41

1 SAFETY INSTRUCTIONS 6 OPTISYS CL 1100www.krohne.com 01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en1.2.2 DisclaimerThe manufacturer will not be li

Page 42

SAFETY INSTRUCTIONS 17OPTISYS CL 1100www.krohne.com01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en1.2.5 Warnings and symbols usedSafety warnings ar

Page 43

2 DEVICE DESCRIPTION 8 OPTISYS CL 1100www.krohne.com 01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en2.1 Scope of deliveryOptional accessories• Flow

Page 44 - KROHNE product overview

DEVICE DESCRIPTION 29OPTISYS CL 1100www.krohne.com01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en2.2 Device description of the systemThe system for